FEATURES
- 1 to 3 minute conditioning times
- Precise control
- Stainless steel construction
- Sample timer
BENEFITS
- Quick on-line color decisions
- Attains specification (even when
considering control, uniformity and
accuracy)
- Durable
- Ease of use
|
The textile conditioning unit is
specifically designed to rapidly bring
textile samples to the standard
condition of 21°C and 65% RH.
Depending on the weight of the
sample, this system will bring
samples to equilibrium within 1 to
3 minutes.
PGC’s humidity generating systems
are unique. By utilizing the “two
temperature” method of control, PGC
is able to create an unrivaled stable
environment (see chart below).
PGC considers all three parameters
necessary for attaining a specific
specification (control constancy,
chamber uniformity, sensor accuracy),
thus providing the specified
tolerance of ±1°C and ±2% RH is truly
attainable. Most of our competition
will control to ±2% RH, but sensor
accuracy and chamber uniformity
must be added to this figure.
| RH Variables |
PGC, Inc. |
Competition |
| Total |
± 2.0% RH |
± 6.0% RH |
| Control Constancy |
± 0.5% RH |
± 2.0% RH |
| Chamber Uniformity |
± 0.5% RH |
± 2.0% RH |
| Sensor Accuracy |
± 1.0% RH |
± 2.0% RH |
|
This system is constructed with durable, corrosion-resistant stainless steel. Low watt density heaters and
various safety cutouts are included.
Unit is available with two 24” D65
fluorescent bulbs. These will be
mounted externally on top of this unit to shine through a lens on to the sample drawers.
A highly precise, digital hygrometer
(temperature compensated) provides
state-of-the-art accuracy while a userfriendly
SmartPad provides optimum
control. Our controller also features
60-segment programming, high/low
(or deviation) alarms, and a serial
communications port.
|